GOM-804 & GOM-805 User Manual
126
Source Commands
SOURce:DRY
Set
Query
Description
Sets or returns the dry circuit test mode.
Syntax
Query Syntax
SOURce:DRY {<NR1> | {OFF|ON}
SOURce:DRY?
Parameter/
Return parameter
<NR1>
0:OFF.
1:ON.
OFF
Turn dry circuit test mode off.
ON
Turn dry circuit test mode on.
Example
SOUR:DRY On
Turns the dry circuit test mode on.
SOURce:DRIVe
Set
Query
Description
Sets or returns the drive mode.
Syntax
Query Syntax
SOURce:DRIVe <NR1>
SOURce:DRIVe?
Parameter/
Return parameter
<NR1>
1: the DC+ mode.
2: the DC- mode.
3: the PULSE mode.
4: the PWM mode.
5: the ZERO mode.
Example
SOURce:DRIVe 3
Sets the drive mode to pulse.