EasyManua.ls Logo

GW Instek GOM-805 - Calculate:compare:limit:result; Calculate:compare:limit:upper; Calculate:compare:math:data

GW Instek GOM-805
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND OVERVIEW
115
CALCulate:COMPare:LIMit:RESult
Query
Description
Returns the compare function test result.
Query Syntax
CALCulate:COMPare:LIMit:RESult?
Return parameter
<NR1>
0: LO
1: IN
2: HI
Example
BINN:LIMit:RES?
>2
Indicates that the test result is HI.
CALCulate:COMPare:LIMit:UPPer
Set
Query
Description
Sets or returns the upper limit value for the compare
function.
Syntax
Query Syntax
CALCulate:COMPare:LIMit:UPPer {<NRf>[,<String>]}
CALCulate:COMPare:LIMit:UPPer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter
<NR3>
000.0000~999.9999E± X
Example
CALC:COMP:LIM:UPP 0.95,kohm
Sets the upper limit value to 0.95kΩ.
CALC:COMP:LIM:UPP?
>0.9500E+3
Returns the upper limit as 0.95kΩ.
CALCulate:COMPare:MATH:DATa
Query
Description
Returns the deviation value for the compare function.
Query Syntax
CALCulate:COMPare:MATH:DATa?
Return parameter
<NR3>
±0.0000~9.9999E± X.
Example
CALC:COMP:MATH:DAT?
>+0.3658E+2
Returns the deviation as 36.58%.