COMMAND OVERVIEW
117
Example
CALC:SCAN:CHAN 5
Sets the channel to 5.
CALCulate:SCAN:DELay
Set
Query
Description
Sets or returns the interval delay for the scan function.
Syntax
Query Syntax
CALCulate:SCAN:DELay <NR1>
CALCulate:SCAN:DELay?
Parameter/
Return parameter
<NR1>
400~30000
Unit:ms
Example
CALC:SCAN:DEL 500
Sets interval delay of the scan to 500ms.
CALCulate:SCAN:LIMit:LOWer
Set
Query
Description
Sets or returns the lower limit value for the scan function.
Syntax
Query Syntax
CALCulate:SCAN:LIMit:LOWer {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:LOWer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter
<NR3>
000.0000~999.9999E± X
Example
CALC:SCAN:LIM:LOW 0.123,maohm
Sets the lower limit value to 0.123MΩ.
CALC:SCAN:LIM:LOW?
>0.1230E+6
Returns the lower limit as 0.123MΩ.
CALCulate:SCAN:LIMit:MODE
Set
Query
Description
Sets or returns the scan function compare mode.
Syntax
Query Syntax
CALCulate:SCAN:LIMit:MODE {ABS|DPER}
CALCulate:SCAN:LIMit:MODE?
Parameter/
Return parameter
ABS
The test results are judged from
absolute values.