COMMAND OVERVIEW
129
Example
SYST:BRIG 4
Turns the brightness level to 4.
SYSTem:ERRor
Query
Description
Returns the current system error, if any.
Query Syntax
SYSTem:ERRor?
Return parameter
<String>
Error number,”Error message”
Example
SYST:ERR?
>0,No error.
Indicates that there is no error message.
SYSTem:HANDler
Set
Query
Description
Sets or returns the handler state.
Syntax
Query Syntax
SYSTem:HANDler {CLEAR | HOLD}
SYSTem:HANDler?
Parameter/
Return parameter
Clear
It clears the last result before executing
measurement.
HOLD
It holds the test result and changes
when a different result appears.
Example
SYST:HAND HOLD
Sets the test result to the hold state.
SYSTem:KEYClick:BEEPer
Set
Query
Description
Sets or returns the keyclick beeper state.
Syntax
Query Syntax
SYSTem:KEYClick:BEEPer <NR1> | {OFF | ON}
SYSTem:KEYClick:BEEPer?
Parameter/
Return parameter
<NR1>
0:OFF.
1:ON.
OFF
Turn the keyclick beeper off.
ON
Turn the keyclick beeper on.
Example
SYST:KEYC:BEEP OFF
Sets the keyclick beeper off.