GW Instek GOM-805 - Trigger Commands; Read; Measure<X; Show

GW Instek GOM-805
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GOM-804 & GOM-805 User Manual
138
Trigger Commands
READ
Query
Description
Returns the measurement value.
Query Syntax
READ?
Return parameter
<NR3>
± 0.0000~5.1000E± X
Example
READ?
>+2.2012E+0
Returns the measurement.
MEASure<X>
Query
Description
Returns the results of the selected channel in the scan
mode, including HI/LO/IN and value.
Query Syntax
MEASure<X>?
Parameter
<X>
Channel 1~100
Return parameter
0|1|2,<NR3>
0:LO
1:IN
2:HI
<NR3>: Measurement result.
Example
MEAS1?
>1,+0.9978E+1
Returns channel 1 as 9.978.
SHOW
Query
Description
Returns the judgments of all (up to 100) channels in the
scan mode.
Query Syntax
SHOW?
Return parameter
<String>
100 characters
0:LO
1:IN
2:HI
_:Channel not active
www.GlobalTestSupply.com