COMMAND OVERVIEW
123
Parameter/
Return parameter
OHM
OHM MODE
COMP
COMP MODE
BIN
BIN MODE
TC
TC MODE
TCONV
TCONV MODE
SCAN
SCAN MODE
DIODE
DIODE MODE
Example
SENS:FUNC OHM
Sets ohm mode on.
SENSe:RANGe
Set
Query
Description
Sets or returns the range of the present function.
Syntax
Query Syntax
SENSe:RANGe <NRf>
SENSe:RANGe?
Parameter
<NRf>
5E-2 ~ 5E+6
Return parameter
<NR3>
5E-2 ~ 5E+6
Example
SENS:RANG 0.05
Sets range to 50mΩ.
SENS:RANG?
>5.0000E-2
Returns the range as 50mΩ.
SENSe:SPEed
Set
Query
Description
Sets or returns the measurement speed.
Syntax
Query Syntax
SENSe:SPEed {SLOW|FAST}
SENSe:SPEed?
Parameter/
Return parameter
SLOW
Measurement speed is slow.
FAST
Measurement speed is fast.
Example
SENS:SPE FAST
Sets measurement speed to the fast rate.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com