COMMAND OVERVIEW
131
Example
SYST:MDEL:DAT 1.105
Sets the delay time of measure is 1.1s.
SYST:MDEL:DAT?
>001.100
Returns the measurement delay as 1.1s.
SYSTem:MDELay:STATe
Set
Query
Description
Sets or returns the measurement delay function state.
Syntax
Query Syntax
SYSTem:MDELay:STATe <NR1> | {OFF|ON}
SYSTem:MDELay:STATe?
Parameter/
Return parameter
<NR1>
0:OFF.
1:ON.
OFF
Turn the measurement delay off.
ON
Turn the measurement delay on.
Example
SYST:MDEL:STAT OFF
Turns the measurement delay function off.
SYSTem:PWM:ON
Set
Query
Description
Sets or returns the duty ON period for the PWM drive
mode.
Note
PWM drive mode is only available for the GOM-805.
Syntax
Query Syntax
SYSTem:PWM:ON <NR1>
SYSTem:PWM:ON?
Parameter/
Return parameter
<NR1>
3~99
Unit: time units. For 60Hz LF, each
unit is equal 16.6ms. For 50Hz LF,
each unit is equal to 20.0ms.
Example
SYST:PWM:ON 5
Sets the duty ON time to 5 adc units.