MEASUREMENT
53
1. Select the
Temperature
Compensation
mode
Press
TC
to access the Temperature Compensation
function.
The temperature-compensated resistance measurement
will appear on the display.
Temperature
compensation
function indicator
Ambient
temperature
source
Extrapolated resistance
measurement at the desired
(“correct”) temperature
Correct Temperature,
Temperature Coefficient settings
Ambient
temperature
2. Ambient
Temperature
The ambient temperature can be either measured with
the PT-100 sensor or be set manually.
If using the PT-100 sensor the Ambient temperature
setting should be turned off. If the PT-100 probe is not
used, then the ambient temperature needs to be manually
set.
From the bottom menu, go to Meas. Setup > Ambient
Temperature and set the ambient temperature.
See page 66 for setting details.
Range
Off, -50.0 ºC ~ 399.9ºC