2. Initial
Resistance,
Initial
Temperature and
Constant settings
Use the arrows keys to go to Initial Resistance, Initial
Temperature or Constant (inferred initial resistance
temperature) and press Enter.
Use the left and right arrow keys to select a digit and use
the up and down arrow keys to edit the digit. Press Enter
to confirm the edit.
Initial Resistance, Initial Temperature
and Constant settings
Move
and edit
Select and
confirm
Enter
Initial Resistance
000.0001~999.9999 mΩ, Ω, kΩ,
MΩ
Initial Temperature
-50.0 ~ +399.9
Constant
000.0~999.9
3. Display mode
Use the arrow keys to go to Disp. Press Enter to toggle
between the T and T modes.
Disp setting
Move
Toggle
Enter
T displays the extrapolated temperature at the measured
resistance of the DUT.
T displays the difference from the extrapolated
temperature at the measured resistance of the DUT and
the ambient temperature. Please refer to page 56 for
further details.
3. Temperature
compensation
connection.
Sensor Connection:
REMOVE INPUTS BEFORE OPENING.
TO AVOID SHOCK,
WARNING
RS232
SER.NO. LABEL
100 240V , 50 60Hz
25VA MAX
AC
HANDLER / SCAN / EXT I/O
TC SENSOR
GPIB
PT-100 temperature
sensor
www.GlobalTestSupply.com