GPP Series User Manual
142
Example
:DELAy2:STATe:GENE 01P
The automatically generated state will repeat by the
sequence of Off and the On for CH2.
Command
:DELAy[1|2]:STATe: GENErate?
Function
Queries the currently used pattern of the
automatically generated state.
Example
:DELAy2:STATe:GENE?
Queries the currently used pattern of the
automatically generated state for CH2.
Command
:DELAy[1|2]:STOP
{NONE|<V|>V|=V|<C|>C|=C|<P|>P|=P}[,<value>]
Function
Sets the stop condition of delay.
Description
NONE: Sets stop condition type as “NONE”.
<V, >V,=V: Sets stop condition type as “<voltage”
“>voltage” =voltage”.
<C,>C, =C: Sets stop condition type as “<current”
“>current” =current”.
<P,>P, =P: Sets stop condition type as “<power”
“>power” =power”.
<value> It is used to set voltage, current or power
value of stop condition, which range from 0 to the
maximum voltage/current/power value of the
active channel.
Example
:DELAy2:STOP >V,8
Sets the stop condition of delay as >8V for CH2.
Command
:DELAy[1|2]:STOP?
Function
Queries the stop condition of the current delay.

Table of Contents