REMOTE CONTROL
127
Function
Queries the current limit state. Returns 0 if the
current limit has not been reached, returns 1 if the
current limit has been reached.
Example
:SOURce2:CURRent:STATe?
Return the current limit state of CH2.
Note: At output off or Load Mode,The return value of is 0.
Command
:SOURce[1|2|3|4]:VOLTage < NRf >
Function
Sets the output voltage amplitude.
Description
<NRf>
0.000-MAX
Example
:SOURce2:VOLTage 5.321
Sets the output voltage to 5.321V for CH2.
Command
VSET<X>:<NR2>
Function
Sets the Voltage level.
Description
X 1= CH1, 2= CH2, (4323: 3 = CH3, 4= CH4)
<NR2>Decimal number, range 0-32.000V
Example
VSET1:20.345
Sets the CH1 voltage to 20.345V.
Command
:SOURce[1|2|3|4]:VOLTage?
Function
Queries the output voltage setting.
Example
:SOURce2:VOLTage?
Queries the output voltage setting on CH2.
Command
VSET<X>?
Function
Returns the output voltage setting.