SYSTEM SETTINGS
91
Firmware Upgrading
When to
Upgrade
Firmware
When system is failure, request by customer
GW Instek. When the system fails, firmware
can requested by GW Instek customers.
Upgrade
Requirement
Firmware file
Supplied by GW Instek
Flash drive
USB2.0/USB3.0FAT32
file system16G MAX
Operation
After plugging in flash drive, the corresponding
symbol appears in the status bar.
Press System button;
Press F3 (Setting) button;
Press F4 (Upgrade) button);
Rotate the scroll wheel to select corresponding
file;
Press the F4 (Recall) button and start to
upgrade;
The procedure is done when the message,
upgrading successful, shows up.