GPP Series User Manual
30
Load Connection
Recommended
Cables
Model
Specification
Usage
GTL-
104A
10A
Front panel terminal
GTL-105A
3A
Sense1326 only
Front panel
wiring
Use the GTL-104A cables for
the front panel source
connections.
USB Type A only
(Greater than 4A)
Use the GTL-105A cables for
the sense connections.
Caution
For safety considerations, please keep in mind
that the wiring must be equivalent to the wiring
on the front terminals.
Wire Gauge
Load wires must have enough current capacity
to minimize cable loss and load line impedance.
Voltage drop across a wire should not excess
0.5V. The following list is the wire current
rating at 450A/cm2.
Wire Size(AWG)
Maximum Current (A)
20
2.5
18
4
16
6
14
10
12
16