Command
:MONItor[1|2|3|4]:VOLTage:CONDition{<V|>V
|=V|NONE},{AND|OR|NONE}
Function
Queries the voltage monitor condition of monitor
(active channel).
Description
{<V|>V|NONE} Sets voltage monitor condition as
“ <V”(<voltage),“ >V”(>voltage) , “=V”(=voltage)or
“ NONE”(none-monitored voltage)
{AND|OR|NONE} Practical monitor condition is
the logical composition of voltage, current and
power. ,”AND ” (and), “OR” (or), “ NONE”(none-
logical pattern)
Example
:MONItor2:VOLTage:CONDition <V,AND
Sets the CH2 voltage monitor condition of monitor
(active channel) as <voltage,and”.
Command
:MONItor[1|2|3|4]:VOLTage:CONDition?
Function
Queries the voltage monitor condition for the
current monitor (active channel).
Example
:MONItor2:VOLTage:CONDition?
Queries the CH2 voltage monitor condition for the
current monitor.
Command
:MONItor[1|2|3|4]:VOLTage[:VALue]{<value>|
MINimum|MAXimum}
Function
Sets the voltage monitor value of monitor (active
channel)
Description
<value) ranges from 0 to the maximum voltage
value of the active channel.
Example
:MONItor2:VOLTage 5
Sets the CH2 voltage monitor value of monitor as
5V (active channel).