LCR-1000 Series
16
Suggestion
For components with low impedance value Z
(high value capacitors and low value inductors)
use the series equivalent circuit,
For components with large impedance Z (low
value capacitors and high value inductors) use
parallel equivalent circuits.
The equivalent circuit is determined according to
the actual use of the component. For example, for
a capacitor, a series equivalent circuit is used for
power filtering, and a parallel equivalent circuit
is used for an LC oscillator circuit.
Test Frequency
LCR-1100: 50Hz, 100Hz, 120Hz, 1kHz, 2kHz,
10kHz, 50kHz and 100kHz
LCR-1010: 50Hz, 100Hz, 120Hz, 1kHz, 2kHz,
10kHz
Accuracy: 0.02%
Test Level
0.3, 0.7 and 1.0 Vrms
Accuracy: 10%
Output Impedance
100Ω
Accuracy: 5%
Range
Auto, Hold range. Total 8 Ranges.