GW Instek LCR-1100 - Measurement Signal Level (LEVEL); Measurement Range (RANGE)

GW Instek LCR-1100
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MEASURE
25
LEVEL
Measurement signal level.
Level Accuracy: 10%
3 levels: 0.3Vrms, 0.7Vrms, 1.0Vrms
RANGE
There are 8 ranges, including: 10Ω, 100Ω, 300Ω, 1kΩ, 3kΩ, 10kΩ,
30kΩ, 100kΩ.
AUTO
Automatically selects the range.
The instrument will select the
appropriate range according to the
impedance|Z| of the device under test.
HOLD
Locks on the current range
The fastest test speed.
INCR +
Locks on the current range and
increases the range number.
DECL -
Lock on current range and decreases
the range number.
Table
Range and
corresponding
test range
Range
number
Range
definition
Impedance measurement
range
7
10
0 10.5
6
100
10.2 ~ 320
5
300
300 ~ 990
4
1k
950 3.2k
3
3k
3k 9.9k
2
10k
9.5k 32k
1
30k
3k ~ 99k
0
100k
95k