LCR-1000 Series
32
OPEN/SHORT
Prior to measurement, the user needs to correct the fixture to
eliminate stray capacitance and series impedance that may be
generated by the fixture.
When the instrument is turned on, in order to achieve the accuracy
of the technical indicators, please perform open circuit test and
short circuit test before measuring after warming up the
instrument.
When replacing the test fixture or test cable, please reset the open
and short test again.
When the temperature changes greatly, please perform open and
short test in time.
On the [MEAS DISPLAY] page, press the [OPEN SHORT] key to
enter the [OPEN SHORT] page.
This page offers two options:
OPEN TEST - open circuit correction
SHORT TEST - short circuit correction