EasyManua.ls Logo

GW Instek LCR-1100 - Accuracy for C and D Measurements

GW Instek LCR-1100
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SPECIFICATIONS
71
Accuracy C and D
50Hz/60Hz/100Hz/120Hz
Range
Display range
Accuracy Ce
Accuracy De
20mF
5.0000mF - 50.000mF
2.0%+5 counts
0.0200
5mF
500.0F - 4.9999mF
1.0%+3 counts
0.0100
500F
50.00F - 499.99F
0.5%+2 counts
0.0050
50F
5.000F 49.999F
0.35%+2 counts
0.0020
5F
500.0nF 4.9999F
0.2%+2 counts
0.0020
500nF
50.00nF 499.99nF
0.2%+2 counts
0.0020
50nF
5.000nF 49.999nF
0.6%+3 counts
0.0030
5nF
500pF 4.9999nF
2.0%+5 counts
1kHz/2kHz
Range
Display range
Accuracy Ce
Accuracy De
1mF
500.0F 4.9999mF
1.0%+5 counts
0.0500
500F
50.00F 499.99F
0.3%+3 counts
0.0300
50F
5.000F 49.999F
0.2%+2 counts
0.0030
5F
500.0nF 4.9999F
0.2%+2 counts
0.0020
500nF
50.00nF 499.99nF
0.2%+2 counts
0.0020
50nF
5.000nF 49.999nF
0.2%+2 counts
0.0020
5nF
500.0pF 4.9999nF
0.3%+3 counts
0.0030
500pF
50.0pF 499.9pF
0.65%+5 counts
10kHz
Range
Display range
Accuracy Ce
Accuracy De
50F
5.000F 49.999F
1.0%+3 counts
0.0100
5F
500.0nF 4.9999F
0.3%+2 counts
0.0030
500nF
50.00nF 499.99nF
0.2%+2 counts
0.0020
50nF
5.000nF 49.999nF
0.2%+2 counts
0.0020
5nF
500.0pF 4.9999nF
0.2%+2 counts
0.0020
500pF
50.00pF 499.9pF
0.3%+3 counts
0.0030
50pF
5.00pF 49.99pF
1.2%+5 counts
50kHz/100kHz
Range
Display range
Accuracy Ce
Accuracy De
5F
500.0nF 4.9999F
3.0%+10 counts
0.0300
500nF
50.00nF 499.99nF
0.3%+5 counts
0.0030
50nF
5.000nF 49.999nF
0.2%+2 counts
0.0020
5nF
500.0pF 4.9999nF
0.2%+2 counts
0.0020
500pF
50.00pF 499.99pF
0.3%+2 counts
0.0020
50pF
5.000pF 49.999pF
1.0%+5 counts
0.0100
5pF
1.000pF 4.999pF
3%+10 counts

Related product manuals