LCR-6000 Series User Manual 
23 
3.4 Warm‐up Time 
LCR-6000 Series is ready to be used as soon as the 
power-up sequence has completed. However,  to achieve 
the specification accuracy, first warm up the instrument 
for 30 minutes. 
3.5 Connect to the Device Under Test 
(DUT) 
The LCR-6000 Series uses a four terminal measurement 
configuration that provides easy, accurate and stable 
measurements and avoids mutual inductance, interference 
from measurement signals, noise and other factors 
inherent with other types of connections. 
Figure 3-3  Connect to DUT 
L Force L Sense H Sense H Force
- DUT +
Do not apply DC voltage or current to the 
UNKNOWN terminals. Applying a DC voltage or 
current may lead to device failure. Connect the 
measurement sample (DUT) to the test port
 
(or 
the test fixture, cables, etc. after the DUT has 
been  completely discharged.