Remote Control
92
Figure 10-2 PC - LCR-6000 Series connection uses a null modem
connection
LCR PC
RxDPin2 RxD Pin2
GNDPin5 GND Pin5
TxD Pin3TxDPin3
Make sure the controller you connect to LCR-6000
Series also uses these settings. The RS-232 interface
transfers data using:
8 data bits, 1 stop bit, no parity.
10.2 To Select Baud Rate
Before you can control the LCR-6000 Series by issuing
RS-232 commands from the built-in RS-232 controller
connected via its DB-9 connector, you have to configure
the RS-232 baud rate.
The LCR-6000 Series built-in RS-232 interface uses the
SCPI language. The RS-232 configuration is as follows:
Data bits: 8-bit
Stop bits: 1-bit
Parity: none
To set up the baud rate
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the [BAUD] field.
Step 4. Use the soft-keys to select the baud rate.
Soft-key
Function
1200
Sets the baud rate to 1200
9600
Sets the baud rate to 9600
38400
Sets the baud rate to 38400
57600
Sets the baud rate to 57600
115200
Sets the baud rate to 115200