EasyManua.ls Logo

GW Instek LCR-6100 - Measure Key (Continued); [LIST MEAS] Page

GW Instek LCR-6100
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measure Key
28
Table 4-5 Effective measurement range for the impedance range
when in HOLD state.
Range
No.
Impedance
range
Effective measurement range
8
10Ω
0 - 10Ω
7
30Ω
10Ω-100Ω
6
100Ω
100Ω-316Ω
5
300Ω
316Ω-1kΩ
4
1kΩ
1kΩ-3.16kΩ
3
3kΩ
3.16kΩ-10kΩ
2
10kΩ
10kΩ-31.6kΩ
1
30kΩ
31.6kΩ-100kΩ ()
0
100kΩ
100kΩ- (This range is available only
when the test frequency < 20kHz.)
Table 4-6 DCR FUNC effective measurement range for the
impedance range when in HOLD state.
When the measured value exceeds 150% of the full
scale for the selected range, “OVERLOAD” will be
displayed.
Range
No.
Impedance
range
Effective measurement
range
Overload
8
1Ω
0Ω 3.1Ω
4.65Ω
7
10Ω
2.8Ω 99Ω
148.5Ω
6
100Ω
90Ω 312Ω
468Ω
5
300Ω
280Ω 990Ω
1.485kΩ
4
1kΩ
900Ω 3.1kΩ
4.65kΩ
3
3kΩ
2.8kΩ 9.9kΩ
14.85kΩ
2
10kΩ
9kΩ 31kΩ
46.5kΩ
1
30kΩ
28kΩ 99kΩ
148.5kΩ
0
100kΩ
90kΩ 312kΩ
100MΩ
DCR accuracy: ±0.05% with effective measurement.
range

Table of Contents

Related product manuals