Measure Key 
48 
There are 4 available trigger modes in the [LIST MEAS] 
page. 
Trigger by the meter’s internal trigger 
source. 
Trigger by pressing the trigger button. 
Trigger by sending a signal to the 
trigger pin on the handler interface. 
Trigger by the RS-232 port. 
4.4.3 Range Mode [RANGE] 
Displays the selected measurement range for each 
measurement step; this field is here only for display, its 
content can’t be changed here. If you need to change the 
measurement range; it can be done in the  
[MEAS DISPLAY] or [MEAS SETUP] function. 
4.4.4 Measurement log [LOG] 
The LCR-6000 has an internal data buffer that record up 
to 10000 measurement readings. These readings can be 
saved to an external USB drive in a .csv file format. 
These readings can then be opened on a PC using 
software such as Windows Excel. When the LOG 
function is active, the display is fixed to the [MEAS 
DISPLAY] screen. The [FUNC], [FREQ] and [LEVEL] 
settings are also fixed and cannot be changed. 
  Procedure for setting Log 
Step 1.  Press the [LIST MEAS] soft-key. 
Step 2.  Use the cursor keys to select the [LOG] status 
field.  
Step 3.  Use the [START LOG] soft-keys to start a 
new log. 
To start a new measurement log. 
Step 4.  After a new log has started, the measurement 
readings will be recorded into the meter’s 
internal buffer. The options in the following 
table are available during this recording 
process: