Setup Key
54
5.1.1 Source Output Impedance [SRC RES]
The Source output impedance can be set to 30Ω, 50Ω or
100Ω.
If you use LCR-6000 Series to test a smaller inductor,
please use 30Ω.
If you need to compare test results with Keysight
E4980A, select 100Ω.
Procedure for setting source output impedance
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select the [SRC RES]
field.
Step 3. Use the soft-keys to set the source output
impedance.
5.1.2 Averaging Factor [AVG]
The averaging feature allows you to obtain moving
average values of successive measurement results.
You can specify the averaging factor from integer 1 to
integer 256.
To set up the averaging factor
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select [AVG] field.
Step 3. Use the soft-keys or number keys to enter
averaging factor.
Increments the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
Decrements the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.