If the Result setting it set to Auto, the meter will 
automatically send out the measurement results each time 
a test is finished; this setting is convenient especially 
when the meter is working with a sorting machine. The 
meter will start a test after receiving the trigger signal and 
then returns the test result to the sorting machine 
without the need to receive a ‘fetch?’ command from 
either the sorting machine or the control PC. 
To set up the Result feature 
Step 1.  Press the [Measure] or [Setup] key. 
Step 2.  Press the [SYSTEM] bottom soft-key.  
Step 3.  Use the cursor key to select the [RESULT] 
field.  
Step 4.  Use the soft-keys to set.  
The data buffer setting sets the maximum number of 
record counts that the internal buffers can store; Settable 
range 1~10000. 
To set up the Data Buffer feature 
Step  1.  Press the [Measure] or [Setup] key. 
Step  2.  Press the [SYSTEM] bottom soft-key.  
Step  3.  Use the cursor key to select the 
[DATA BUFFER] field.
  
Step  4.  Enter the number of sets using the entry keys 
or use the soft-key to select the maximum 
number of sets.