Step 4.  Press the [MEAS OPEN] soft-key. A dialog 
message, “Open-circuit the test terminals”, will 
appear. 
Step 5.  Connect the test fixture to the BNC terminals. 
Do not connect the DUT to the test fixture. 
Step 6.  Press the [OK] soft-key. The LCR-6000 
measures the open admittances (capacitance 
and conductance) at this frequency point. 
    During the measurement, an “OPEN 
measurement in progress” dialog message will 
be shown on the display. 
    When the measurement has finished, 
“Correction finished”, will be displayed. 
During the measurement, you can press the 
[ABORT] soft-key to abort open correction. 
  To specify frequency points and perform short 
correction 
Step 1.  Press the [MEAS  SHORT]  soft  key,  a  
dialog  message, “Short-circuit  the  test 
terminals”, will be displayed. 
Step 2.  Connect the test fixture to the BNC terminals 
and short-circuit the test terminals. 
Step 3.  Press the [OK] soft-key. The LCR-6000 
measures the short impedances (inductances 
and resistance) at this frequency point. 
    During the measurement, a “SHORT 
measurement in progress” dialog message will 
be shown on the display. 
    When the measurement has finished, 
“Correction finished” is displayed on the 
screen. During the measurement, you can press 
the [ABORT] soft-key to abort short 
correction.