External voltage control of the current output is 
accomplished using the analog control 
connector on the rear panel. There are two 
external voltage control ranges, 0~5V and 
0~10V, depending on the F-97 configuration. 
See page 108 for details. 
For 0~10V: 
Output current = full scale current x (external 
voltage/10) 
For 0~5V: 
Output current = full scale current x (external 
voltage/5) 
When connecting the external voltage source to 
the connectors, use shielded or twisted paired 
wiring.