APPENDIX
189
Analog Programming and Monitoring
Model
PSU
6-200
12.5-120
20-76
40-38
60-25
External voltage control
output voltage
Accuracy and linearity: ±0.5% of rated output
voltage.
External voltage control
output current
Accuracy and linearity: ±1% of rated output
current.
External resistor control
output voltage
Accuracy and linearity: ±1% of rated output
voltage.
External resistor control
output current
Accuracy and linearity: ±1.5% of rated output
current.
Output voltage monitor
Accuracy: ±1%
Output current monitor
Accuracy: ±1%
Shutdown control
Turns the output off with a LOW (0V to 0.5V)
or short-circuit.
Output on/off control
Possible logic selections:
Turn the output on using a LOW (0V to 0.5V)
or short-circuit, turn the output off using a
HIGH (4.5V to 5V) or open-circuit.
Turn the output on using a HIGH (4.5V to 5V)
or open-circuit, turn the output off using a
LOW (0V to 0.5V) or short-circuit.
Alarm clear control
Clear alarms with a LOW (0V to 0.5V) or short-
circuit.
CV/CC/ALM/PWR
ON/OUT ON indicator
Photocoupler open collector output;
Maximum voltage 30V, maximum sink current
8mA.
Trigger out
Maximum low level output = 0.8V; minimum
high level output = 2V; Maximum source
current = 8mA.
Trigger in
Maximum low level input voltage = 0.8V;
minimum high level input votage = 2.0V,
Maximum sink current = 8mA.