CV and CC Slew Rate Priority are disabled for V-I
mode (F-03) when using external voltage control.
See the normal function settings on page 99.
The isolated analog connector can also be used
to monitor the current (IMON) or voltage
(VMON).
For the isolated voltage option, an output of
0~5V or 0~10V represents the voltage or current
output of 0~ rated current/voltage output.
IMON = (current output/full scale) × (5 or
10)
VMON = (voltage output/full scale) × (5 or
10)
For the isolated current option, an output of
4~20mA represents the voltage or current
output of 0~ rated current/voltage output.
IMON = ((current output/full scale) ×
16mA) + 4mA
VMON = ((voltage output/full scale) ×
16mA) + 4mA