PSU Series User Manual 
168 
Multiple Unit Connection Using Local RS485 Bus 
The PSU power supplies can have up to 31 units daisy-chained 
together using the 8 pin connectors (IN OUT ports) on the rear 
panel. The first unit in the chain is remotely connected to a PC 
using RS232 or RS485. Each subsequent unit is daisy-chained to the 
next using an RS485 local bus. The OUT port on the last terminal 
must be terminated by the end terminal connector. 
Only the TDK GENESYS legacy commands can be used when 
using multiple units over the local RS485 bus, SCPI commands 
cannot be used. When using the TDK commands, each unit is 
assigned a unique address and can then be individually controlled 
from the host PC. When using SCPI commands via RS232/RS485, 
remote commands can only be used to control the master unit. 
 
1.  Connect the first unit’s IN port to a PC via 
RS232 or RS485. 
  Use the serial cables supplied in the PSU-
232 or PSU-485 connection kit.  
2.  Connect the OUT port on the first unit to the IN 
port of the second unit using the serial link 
cable supplied in the PSU-232 or PSU-485 
connection kit.  
3.  Connect all the remaining units in the same 
fashion until all the units have been daisy-
chained together.