GETTING STARTED 
27 
Alarms 
The PSU power supplies have a number of protection features. 
When one of the protection alarms are tripped, the ALM icon on 
the display will be lit and the type of alarm that has been tripped 
will be shown on the display. When an alarm has been tripped the 
output will be automatically turned off. For details on how to clear 
an alarm or to set the protection modes, please see page 51. 
 
Over voltage protection (OVP) prevents a high 
voltage from damaging the load. This alarm 
can be set by the user. 
Over current protection prevents high current 
from damaging the load. This alarm can be set 
by the user. 
Under voltage limit. This function sets a 
minimum voltage setting level for the output. 
It can be set by the user. 
Over temperature protection for slave and 
master board. OHP is a hardware protection 
function. Only when the unit has cooled can 
the over temperature protection alarms be 
cleared. 
Master board over temperature protection.  
Slave board over temperature protection.  
Sense alarm. This alarm will detect if the sense 
wires have been connected to the wrong 
polarity. 
Hardware over voltage protection. This is a 
hardware OVP that is fixed at approximately 
120% of the rated voltage output.