ANALOG CONTROL
145
5. Press the Output key. The
voltage can now be
controlled with the isolated
external voltage control.
Output
Note
The input impedance for isolated external voltage
control is 1M.
Use a stable voltage supply for the external voltage
control.
Note
CV and CC Slew Rate Priority are disabled for V-I
mode (F-03) when using external voltage control.
See the normal function settings on page 99.
CAUTION
Ensure the voltage polarity is correct when
connecting the external voltage.
Ensure no more than 10.5V (for the 0~10V setting)
or 5.25V (for the 0~5V setting) are to be input into
the isolated voltage input.