OPERATION
55
4. Use the Current knob to set the F-
03 setting.
Set F-03 to 0 (CV High Speed
Priority) or 2 (CV Slew Rate
Priority).
Current
F-03
0 = CV High Speed Priority
2 = CV Slew Rate Priority
5. Press the Voltage knob to save the
configuration setting. ConF will be
displayed when it is configuring.
Voltage
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
VSR indicator for CV Slew
Rate Priority (F-03=2)
6. If CV Slew Rate Priority was chosen as the
operating mode, set F-04 (Voltage Slew Rate
Up) and the F-05 (Voltage Slew Rate Down)
and save.
F-04 / F-05
0.001V~0.06V/msec (PSU 6-200)
0.001V~0.125V/msec (PSU 12.5-120)
0.001V~0.2V/msec (PSU 20-76)
0.001V~0.4V/msec (PSU 40-38)
0.001V~0.6V/msec (PSU 60-25)
7. Press the Function key again to exit
the configuration settings. The
function key light will turn off.
Function
8. Use the Current knob to set the
current limit (crossover point).
Current

Table of Contents