PSU Series User Manual
76
Configuration settings can be checked on both the
master and slave units by pressing the Function key
and checking F-93.
Only the Master OVP, OCP and UVL settings are
used for protection. Slave protection levels are
disregarded.
OHP works independently for each unit.
Only operate the power supplies in parallel if
the units are configured correctly.
1. Turn on the master and slave units. The slave
unit(s) will show a blank display.
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
2. Operation of all units is controlled
via the master unit. Operation of
the master unit is the same as for a
single unit. See the Basic Operation
chapter.
3. Press the Output key to
begin. The output LED will
become lit.
Only operate the power supplies in parallel if using
units of the same model number.
The panel controls are disabled on slave units,
including the output key. On slave units, only the
Function key can be used to view the current
settings.