Table of contents
Specifications on page 3 Operation on page 18
Product overview on page 6 Calibration and adjustment on page 18
Product components on page 7 Maintenance on page 18
Installation on page 7 Troubleshooting on page 24
User navigation on page 17 Replacement parts and accessories on page 25
Configuration on page 17
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification Details
Dimensions (W x H x D) 32.9 x 34.2 x 17.7 cm (12.9 x 13.5 x 7.0 in.)
Enclosure IP66 per IEC 60529 with the door closed and latched
1
Shipping weight 4.1 kg (9 lb) without bottles; 5.1 kg (11.2 lb) with full bottles
Mounting Wall mount
Protection class III
Pollution degree 3
Electrical installation category I (indoor installation)
Power requirements 12 VDC, 400 mA maximum (supplied by the controller)
Operating temperature 5 to 40 °C (41 to 104 °F)
Operating humidity 0 to 90% non-condensing relative humidity
Storage temperature -40 to 60 °C (-40 to 140 °F)
Fittings Sample line: ¼-in. OD quick-connect fitting for plastic tubing
Drain lines: slip-on fitting for ½-in. ID soft plastic tubing
Indicator lights Analyzer status and measurement cycle
Certifications CE DoC, FCC SDoC, IC, RCM, KC
Warranty 1 year (EU: 2 years)
Table 2 Sample requirements
Specification Details
Pressure 0.31 to 5.17 bar (4.5 to 75 psig) supplied to Y-strainer; 0.1 to 0.34 bar (1.5 to 5 psig) supplied to
analyzer
Flow rate 60 to 200 mL/min through the instrument
Temperature 5 to 40 °C (41 to 104 °F)
Filtration Y-strainer with 40 mesh screen or higher
1
Drain holes closed with plugs for testing requirements.