EasyManua.ls Logo

Hach CL17sc - Specifications

Hach CL17sc
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table of contents
Specifications on page 3 Operation on page 18
Product overview on page 6 Calibration and adjustment on page 18
Product components on page 7 Maintenance on page 18
Installation on page 7 Troubleshooting on page 24
User navigation on page 17 Replacement parts and accessories on page 25
Configuration on page 17
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification Details
Dimensions (W x H x D) 32.9 x 34.2 x 17.7 cm (12.9 x 13.5 x 7.0 in.)
Enclosure IP66 per IEC 60529 with the door closed and latched
1
Shipping weight 4.1 kg (9 lb) without bottles; 5.1 kg (11.2 lb) with full bottles
Mounting Wall mount
Protection class III
Pollution degree 3
Electrical installation category I (indoor installation)
Power requirements 12 VDC, 400 mA maximum (supplied by the controller)
Operating temperature 5 to 40 °C (41 to 104 °F)
Operating humidity 0 to 90% non-condensing relative humidity
Storage temperature -40 to 60 °C (-40 to 140 °F)
Fittings Sample line: ¼-in. OD quick-connect fitting for plastic tubing
Drain lines: slip-on fitting for ½-in. ID soft plastic tubing
Indicator lights Analyzer status and measurement cycle
Certifications CE DoC, FCC SDoC, IC, RCM, KC
Warranty 1 year (EU: 2 years)
Table 2 Sample requirements
Specification Details
Pressure 0.31 to 5.17 bar (4.5 to 75 psig) supplied to Y-strainer; 0.1 to 0.34 bar (1.5 to 5 psig) supplied to
analyzer
Flow rate 60 to 200 mL/min through the instrument
Temperature 5 to 40 °C (41 to 104 °F)
Filtration Y-strainer with 40 mesh screen or higher
1
Drain holes closed with plugs for testing requirements.
English 3

Table of Contents

Related product manuals