24
Standard Operations
5.2.6 PC and printer
The LICO 500 is provided with 2 USB interfaces, which are located
on the back of the instrument (see Figure 2 on page 11). These
interfaces can be used for exporting data and graphics to a printer,
updating data and for data communication to a personal computer.
These interfaces can be used for the connection of a USB stick, an
external USB keyboard or a USB Barcode handset scanner.
Note: A USB hub may be used to connect several accessories at a time.
A USB memory stick is used to upgrade data, see section 6.1.2 on
page 76.
Important Note: A screened USB cable must not be longer than
3m!
1. Press PC & Printer in the Instrument Setup.
A list with information about the connections opens.
5.2.6.1 Printer setup
For reasons of compatibility, the printer language must be
HP PCL 3.
1. Press Printer.
2. Press Setup to display the Printer Setup screen.
Table 3 USB connector
USB Interfaces Description
USB (Type B)
This USB interface is only intended for the LICO 500 to PC connection (with installation of the
HACH Data Trans Software).
USB (Type A) This USB port can be used to connect a printer, a USB memory stick and keyboard.