EasyManuals Logo

Hach ORBISPHERE 51 Series User Manual

Hach ORBISPHERE 51 Series
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
17
Specifications
2.5 Default parameters
The table below indicates the factory default configurations. The instrument has these settings
when started for the first time.
Parameter Default settings Customer settings
Security Enabled
Measurement
Measurement mode Continuous
Measurement interval (TC sensor) 20 seconds (approximately)
Measurement interval (LDO sensor) 4 seconds until 2000 ppb
Data filter Disabled
Medium Liquid
Gas unit type Dissolved
Gas unit ppm-ppb
Liquid Water
Display resolution (TC sensor) XX.XX
Display resolution (LDO sensor) XXX.X
Storage mode Rolling buffer
User offset Disabled
Out of range protection Enabled
Temp unit °C
Pressure unit bar
Calibration
Mode (LDO sensor) Zero calibration
Auto-end (LDO sensor) Enabled
Hold Enabled
Analog outputs
Range 4-20 mA
Outputs Gas measurement
Extended mode Disabled
Characteristics Mono linear mode
Alarm relays Disabled
Thermal cutoff Enabled
Thermal cutoff temp (TC sensor) 65°C
Thermal cutoff temp (LDO sensor) 40°C
Calibration timer Disabled
Service timer Disabled
Buzzer
Screen tap Enabled
Alarm sound Disabled
Display
Minigraph Enabled
Temperature Disabled

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach ORBISPHERE 51 Series and is the answer not in the manual?

Hach ORBISPHERE 51 Series Specifications

General IconGeneral
BrandHach
ModelORBISPHERE 51 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals