4.5.1.1 Cable preparation
W A R N I N G
To conform with security standards and to respect the EMC compliancy of the analyzer, this
procedure must be followed for all user supplied cables before connecting them to the electronic
unit of the analyzer.
Materials required:
• Shielded cable (diameter minimum 4.5 mm - maximum 6 mm) with 2 or 3 wires
depending on function (RS232 = 3 wires, Alarm = 2 wires, etc.)
• Metal cable gland
• Grounding ferrule
• 2 or 3 protective plugs for the exposed wires
• Clamp for the grounding ferrule
• Crimping pliers for the protective plugs
Installation
13