J4, J5, J6 and J7 Terminal number Color
1 Not used
2 Not used
3 Brown
4 White
4.5.3.5 RS232 connection
Refer also to Figure 8 on page 18 and Table 4 on page 19. Wire the connection to J10 as
follows:
J10 Terminal number DB9 Plug DB25 Plug
1 (TXD) RXD: 2 RXD: 2
2 (RXD TXD: 3 TXD: 3
3 (GND) COM: 5 COM: 7
The 8810 configuration is:
• Speed: 9600 baud
• Data: 8 bits
• Stop bit: 2
• Parity: none
On startup the analyzer sends the name and software version:
• ANALYZER 8810 : 00 : 00 : 00 :
• ISE X.XX : 00 : 00 : 00 :
In measurement mode, the analyzer sends the following data:
• Measure HH : MM : SS
• M1 XXXXEXX XX.X°C XXX
Where:
• MX = M1 for channel 1 or M2 for channel 2
• XXXXEXX = concentration measurement value
• XX.X°C = temperature
• XXX = value of the potential
Where appropriate, the analyzer also sends the following messages with a time stamp
(HH : MM : SS):
ACTIONS
Cleaning Chemical cleaning
Stopped Analyzer stopped
Fix-time Time between two cycles
Standby Analyzer in standby
AUTO SYS CAL Automatic calibration
PROCESS SYS CAL Process calibration
MANUAL SYS CAL Manual calibration
Installation
21