Section 1 Specifications
Specifications are subject to change without notice.
Specification Details
Ambient temperature 5 - 40 °C (41 - 104 °F)
Relative humidity 10 to 80%
Operating altitude From 0 to 2,000 m. (6,550 ft.) above sea level
Mains power supply 110/220/240 VAC, 50/60 Hz, ± 10%
Max. consumption 100 VA
Overvoltage category 2 (according to standard EN 61010-1)
EMC requirements EMC Directive 2004/108/EC; Low-voltage Directive 2006/95/EC; EN 61326-1
N O T I C E
This is a Class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
European safety standards EN 61010-1
Compressed air 4 - 7 bar, filtered and dried
Rinse water pressure 0.5 - 6 bar
Reagents 10 liter storage canister (supplied)
Analog outputs Number: 2; 0 - 20 or 4 - 20 linear, logarithmic, or dual programmable
Alarms 2 x Relay (concentration); 1 x Warning or System alarm
Exchangeable sensors ISE, reference electrode, Pt100
Level control Sample; Reagents; Calibration solution; Chemical cleaning
Pumps Peristaltic, micro piston, pulse or volumetric pumps for calibration and conditioning reagents
Tubing Tygon
®
; Polyethylene
Reagent consumption Application specific
Cycle time Programmable up to 999 minutes
Accuracy < ± 2 to 4% (application specific)
Reproducibility < ± 2 to 4% (application specific)
Sample lines 1
Sample temperature 0 - 50 °C (32 - 122 °F)
Sample pressure 0.5 - 6 bar
Sample flow rate 40 - 300 liters/hour
Sample volume/cycle 200 - 1000 mL (adjustable)
Panel mount 743 x 482 x 122 mm (H x W x D); 25 kg
Cabinet (IP 54) 1900 x 600 x 400 mm (H x W x D); < 100 kg
3