EasyManua.ls Logo

Hach POLYMETRON 8810 - Connecting Peripherals

Hach POLYMETRON 8810
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Place the nut and metal seal of the
cable gland onto the shielded cable.
2. Remove 8 cm of the outer cabling to
reveal the braided shielding.
3. Remove all but 12 mm of the
shielding to reveal the interior wires,
then roll the remainder of the shielding
back over the cable.
4. Insert the grounding ferrule over the
shielding.
5. Use the clamp to fix the ferrule
securely to the cable.
6. Remove 5 mm of the external
insulation of the wires.
7. Use the crimping pliers to attach the
protective plugs to the wires.
8. Bring the metal shield up over the
ferrule and lock the two together
making sure the ferrule is not exposed
at the cable end of the shield. Bring up
the nut to cover the shield.
9. Pass the threaded part of the cable
gland over the wires and screw into the
nut on the cable. The cable is now
ready to be connected following the
procedure described in Connecting
peripherals on page 14.
4.5.1.2 Connecting peripherals
W A R N I N G
To conform with security standards and to respect the EMC compliancy of the analyzer, this
procedure must be followed for connecting all peripherals to the electronic unit of the analyzer.
Installation
14

Table of Contents

Other manuals for Hach POLYMETRON 8810

Related product manuals