8.1 Automatic calibration
This option allows the analyzer to perform an automatic calibration at programmed
intervals and with a known calibration solution. Three different kits are available, each
including a 10 liter canister equipped with level detector and strainer:
1. 1 point calibration with a volumetric pump.
2. 2 point calibration with a micro piston pump.
3. 2 point calibration with a pulse pump.
For information regarding the programming and use of this option, refer to the section
entitled Calibration (CALIBRATION option) on page 26.
8.1.1 One point calibration with a volumetric pump
D A N G E R
Electrocution hazard. Always make sure the analyzer is turned off and power is disconnected
before installing this option.
Figure 11 Automatic calibration option
1 Pump 3 Polyethylene tubing 5 Input block (already installed on the
analyzer)
2 Elbow fitting 4 Connection to attach to top of input
block
6 Electric cable
Install the pump as follows (the numbers in brackets refer to Figure 11):
1. Mount the pump (1) in position 4 on the rear of the panel (see Figure 10 on page 37).
2. Connect the tubing (3) between the pump (2) and the input block (5) installed on the
measurement cell cover using the fitting (4).
3. Connect the cable (6) to terminals 1 and 11 of J14 in the electronic unit, and make
sure that jumper X4 is installed between terminals 2 and 3 and a strap placed
between terminals 8 and 10 of J14 (refer also to Figure 8 on page 18 and Table 6
on page 20).
To install the 10 liter canister, refer to Canister kit on page 48.