Option Description
WARNING/R1 YES/NO Define the alarm allocation (see table above).
ALARM 1/2 ON/OFF Activate or deactivate alarms 1 and 2.
ALARM 1/2 HIGH/LOW If activated, define if the trigger is a high or low measurement value.
ALARM 1/2 XXX Enter the threshold value. If the alarm is set to HIGH any measured
value above this threshold will trigger the alarm, or if the alarm is set
to LOW any measured value below this threshold will trigger the
alarm.
6.5 Measurement mode (SEQUENCE option)
Define the measurement mode.
1. Select SEQUENCE with the scrolling keys and press Enter. Use the scrolling keys to
select the required mode.
Option Description
FIX TIME A fixed interval between measurements. The interval is defined in Measurement
cycle parameters (TIMING option) on page 31.
LOOP Continuous measurement. No further input is required.
2. If FIX TIME is selected:
Option Description
Inj. Time PX: YYs Adjust the pump action time:
X: pump number 1 to 4
YY: time in seconds. Adjustable between 0 and 99 seconds.
6.6 Test analyzer and accessories (SERVICE option)
Check various physical options (e.g. pumps) are working correctly. The option to view the
version number of the latest installed software is also available from the list.
1. Select SERVICE with the scrolling keys and press Enter.
2. Select the option to test from the list available and set to ON to start it, check it
functions correctly, and then set back to OFF to stop it.
6.7 Chemical cleaning (CLEANING option)
This option is displayed only if the CLEANING option has been activated.
Define the parameters required for a chemical cleaning of the system.
1. Select CLEANING with the scrolling keys and press Enter.
Option Description
FREQUENCY Enter the number of measurements between each chemical cleaning.
SPRINKLER Define the duration (in seconds) for rinsing the measurement chamber.