Option Description
REAG. INJ Define the duration (in seconds) that the cleaning solution is injected into the
measurement chamber.
RESIDENC. Define the duration (in seconds) that the cleaning solution remains in the
measurement chamber.
6.8 Sample conditioning (SAMPLE COND option)
This option is displayed only if the SAMPLE COND option has been activated.
Define the parameters required for conditioning the sample prior to measurement. It is
possible to define 1 or 2 conditioning reagents.
1. Select SAMPLE COND with the scrolling keys and press Enter.
Option Description
REAGENT 1/2 OFF/ON Set to ON if the conditioning is made using this reagent.
INJ. REAG. 1/2 Define the duration (in seconds) that the reagent is injected into the
measurement chamber.
TIME REAG. 1/2 Define the duration (in seconds) that the reagent remains in the
sample before measurement.
6.9 View calibration details (CAL PARAMETER option)
This option lets you view the details of the primary and last calibrations made by the
analyzer.
1. Select CAL PARAMETER with the scrolling keys and press Enter.
2. Use the scroll keys to select LAST CAL or PRIMARY CAL, then scroll through the
details by pressing Enter.
6.10 Measurement cycle parameters (TIMING option)
Define the parameters for each measurement.
1. Select TIMING with the scrolling keys and press Enter.
Option Description
SAMPLE TIME Define the time (in seconds) the sample is injected into the measuring
chamber.
SPRINKLER N Define the number of rinses required.
SPRINKLER t Define the duration (in seconds) of each rinse.
CYCLE TIME Define the duration (in minutes) of the measurement cycle.
DELAY t Define the time (in seconds) between each injection.
DEVIATION Define the electrode potential.
MEASURE t Define the measurement time. The final measurement is the average of all
measurements taken during this time.