1. Select an unused opening nearest to
the cable connection on the electronic
board. Remove the screw and nut
combination and set aside the screw
for later use.
2. Take the cable, unscrew the
threaded part and remove it.
3. Screw the threaded part into the
electronic unit opening selected in step
1 using the nut removed in step 1 to
secure in place.
4. Pass the rest of the cable through
the gland.
5. Pull the cable slightly with one hand
until the inner shield comes into contact
with the gland. Tighten the nut onto the
gland with the other hand.
6. Tighten the nut with a wrench to
secure in place. It may be necessary to
use a second wrench to hold the top
nut in place.
7. Connect the wires to the correct
terminals.
8. Secure the wires with a cable tie.
9. Cut the cable tie as near as possible
to the wires.
Installation