User configuration
Default value User value Date
SAMPLE COND: sample conditioning
REAGENT 1: XXX Conditioning activated by reagent 1 OFF
REAGENT 2: XXX Conditioning activated by reagent 2 OFF
INJ. REAG. 1: XXXs Duration of reagent 1 injection 10 seconds
INJ. REAG. 2: XXXs Duration of reagent 2 injection 10 seconds
TIME REAG. 1: XXXs Length of time reagent 1 remains in the reactor 20 seconds
TIME REAG. 2: XXXs Length of time reagent 2 remains in the reactor 20 seconds
TIMING
SAMPLE TIME Injection sample time 001 seconds
SPRINKLER N Rinse number 3
SPRINKLER t Rinse time 2 seconds
CYCLE TIME Cycle time 15 minutes
DELAY TIME Stabilization time before measurement 001 seconds
MEASURE t Measurement time 001 seconds
SEQUENCE
FIX TIME/LOOP Measurement sequence LOOP
ALARM
SYS.ALARM ON/OFF System alarm OFF
WARNING/R1 YES/NO Warning alarm on alarm relay 1 NO
ALARM 1 ON/OFF Alarm 1 activation OFF
ALARM 1 HIGH/LOW High or low alarm HIGH
ALARM 1 XXX Threshold value 0
ALARM 2 ON/OFF Alarm 2 activation OFF
ALARM 2 HIGH/LOW High or low alarm HIGH
ALARM 2 XXX Threshold value 0
ANALOG OUT
OUT LINEAR/LOGARITHMIC/DUAL Output current type LINEAR
START XX.XXEXX Beginning of concentration range 00.E00
END XX.XXEXX End of concentration range 10.00E01
START XXX.XmV Beginning of potential range 000.0 mV
END XXX.XmV End of potential range 100 mV
User configuration