Microlab
®
STAR User Manual
610766/02 Page 75 of 225
&DVH L(S
asp
) > L(S
disp
), Control: ASP
The target sequence is repeated to match the length of the source sequence.
Another example for this case: All Tubes to Plate(s). Reloadable plate sequence to load all
tubes to as many plates as needed.
&DVHL(S
asp
)> L(S
disp
), Control: DISP
Source sequence is cut to match length of target sequence; stops when plate is full (prepare
plate for assay)
&DVH L(S
disp
) > L(S
asp
), Control: ASP
Target sequence is cut to match length of source sequence; stops when tubes are processed
(how many samples today?). If L(S
asp
)>96, plate is filled again!
&DVH L(S
disp
) > L(S
asp
), Control: DISP
Source 1
Source 2 Target 1 Target 1