Microlab
®
STAR User Manual
Page 78 of 225 610766/02
Under “DGYDQFHG, pre-rinsing and liquid following can be specified.
Furthermore, sequence manipulations can be made (separately for aspiration and dispense
sequences). If the sequences which this SMART Step is going to work with have been used
within this method in preceding steps, the sequence counters can be reset using the two
check boxes for initial sequence manipulation. If other steps are following this SMART Step,
the status of the sequence which is passed back can be defined by the radion buttons under
“final sequence manipulation“.
Under FKDQQHOVHWWLQJV, the user may specify a channel pattern manually, as well as the
number of the channel which will be used to calibrate (1536-well) carriers.
The HUURUVHWWLQJV dialog allows you to specify an error handling approach:
The choices in case of an error are: abort the method (only abort button on runtime error
dialog); give the choice (in runtime) to abort or to cancel (continue with user-defined error
handling, if programmed, otherwise abort); or use the fixed default error recovery
(recommended) pre-programmed for the SMART Step. In this case two choices can be
made: