Ozone
Press the Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement.
Remove the cuvette.
Add the content of one packet of HI 93757-0
Ozone Reagent. Replace the cap and shake gently
for 20 seconds.
Replace the cuvette into the holder and close the lid.
Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 2 minutes and press Read. When the timer ends the meter will perform the reading.
The instrument displays concentration in mg/L of ozone (chlorine free samples only).
ADDITIONAL MEASUREMENT PROCEDURE
For samples containing chlorine
Select the
Ozone
method using the procedure described
in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample, up to
the mark, and replace the cap.
Place the cuvette into the holder and close the lid.
10 mL