114
Potassium HR
POTASSIUM HIGH RANGE
SPECIFICATIONS
Range 20 to 200 mg/L
Resolution 5 mg/L
Accuracy ±30 mg/L ±7% of reading at 25 °C
Typical EMC ±5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93750A-0 Potassium Reagent 6 drops
HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the CONCENTRATED SAMPLES procedure at page 18.
Select the
Potassium HR
method using the procedure
described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of sample, up to the mark.
Add six drops of HI 93750A-0, replace the cap and swirl
the solution.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83099 and is the answer not in the manual?

Hanna Instruments HI 83099 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83099
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals