Manganese LR
MANGANESE LOW RANGE
SPECIFICATIONS
Range 0 to 300 μg/L
Resolution 1 μg/L
Accuracy ±10 μ/L ±3% of reading at 25 °C
Typical EMC ±1 μg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93748A-0 Ascorbic acid 2 packets
HI 93748B-0 Alkaline-cyanide sol. 0.40 mL
HI 93748C-0 0.1% PAN indicator 2 mL
HI 93703-51 Dispersing Agent 4-6 drops
REAGENT SETS
HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Manganese LR
method using the procedure described in the
Method Selection
section (see
page 12).
Fill one cuvette with 10 mL of deionized water (up to the mark).
Fill a second cuvette with 10 mL of sample (up to the mark).
Add the content of one packet of HI 93748A-0 Ascorbic acid to
each cuvette, replace the caps and shake gently until completely
dissolved.
10 mL
10 mL